A sample is read in a cuvette twice as long as the calibration cuvette. At unchanged concentration in the Beer–Lambert range, what happens to absorbance?
Reveal Answer and Explanation
It doubles.
Absorbance is proportional to optical path length as well as concentration.
T2 · Apply
- From Interference to Molecular Evidence · 3. Keep Absorbance and Concentration Separate
- UV–visible absorption: connect color, conjugation, and concentration · Worked application: use a calibration within its range